Back to top
Send SMS Send Inquiry
Home » Products » Ambient & Work Place Air Monitoring Instruments » Ambient Fine Dust Sampler

Ambient Fine Dust Sampler

Ambient Fine Dust Sampler
Ambient Fine Dust Sampler
Send Inquiry
Product Code : AFDS 2.5 micron
Brand Name : POLLTECH INSTRUMENTS
Product Description

Bestowed with ISO 9001:2008 certification, our enterprise is betrothed to manufacture, export and supply Ambient Fine Dust Sampler. The offered sampler is extensively demanded to measure the concentration of particulate matter in the atmosphere that have size less than 2.5µ. Available in different technical specifications to suit the specific demands of clients, this sampler is manufactured by our ingenious professionals using the finest quality components and pioneering techniques in line with international quality standards. Additionally, clients can purchase this Ambient Fine Dust Sampler from us at the most competitive prices.

Features:

  • Easy to operate

  • Authentic results

  • Longer service life

  • Facilitates quick measurement  

Details:

Model PEM-ADS 2.5µ is an advanced fine dust sampler, which can be used to determine the concentration of particulate matter of size less than 2.5µ in atmosphere. It has been specifically designed to comply with EPA reference method for determination of PM2.5, as published in Federal Register 40 CFR Part 50 Appendix L. (update May 1, 2002). When used without WINS Impactor (2.5µ cut stage), the instrument becomes a standard reference method instrument for particulate matter less than 10µ.

SALIENT FEATURES

  • Complies with US EPA PM2.5 Reference Method
  •  PM 10 & TSP Sampling Options
  • Improved PM10 Sampling Inlet
  • Electronic Flow Regulation to 2%  Accuracy.
  • mP Based Technology
  • Programmable Real Time Clock
  •  Versatile HMI
  •  Comprehensive Data Logging
  • Serial Port for In-Field Data Transfer
  •  Sturdy Construction for Outdoor Use
  • Easy Set-Up and Operation

 

Model PEM-ADS 2.5µ comprises an improved louvered PM-10 sampling Inlet, a WINS impactor, a filter holder for 47mm dia. filter, an orifice assembly for flow measurement, a proportioning valve for flow control and an electronic control module for measurement, control and recording of various parameters.  All these assemblies along with ambient pressure and temperature sensors are housed in an enclosure suitable for outdoor environment.  The Sampling Pump can be provided either in the same enclosure or in a separate housing.

PEM-ADS 2.5µ utilizes an advanced microprocessor-based technology with non-volatile RAM, 4 line LCD display and RS-232 serial interface for computer compatibility. The control module is user friendly, easy to configure and operate and has an intelligent flow and temperature monitoring and regulating systems along with a barometric pressure measurement and recording system.  Flow rate is regulated to âہ¡Ãƒâ€šÃ‚±2%. The temperature monitoring system monitors the temperature of the filter as well as the ambient temperature and the temperature control system ensures that the differential is within ±50C.  The various parameters and measurements can be transferred to a laptop in the field. Optionally the same can be transferred onto a data memory module for off-field transfer to a PC.  Application Software under Windows environment is also provided for enabling data transfer and computation of the final results on the PC.

 

Detailed Technical Specifications

 

US EPA Specifications Compliance

PEM-ADS 2.5µ meets and exceeds USEPA Specifications for  measurement, control and recording of various parameters including   flow rate, Ambient Barometric Pressure and Temperature, filter temperature, temperature differential, power interruption, leak check, automatic start and stop and calibration mode etc.

Sampling Inlet

Improved Louvered PM10 Inlet with sample transport tube

 PM 2.5 Size Separator

EPA PM 2.5 WINS Impactor.

Filter Holder Assembly

Suitable for 47mm dia. PTFE, Nylon, Glass Fiber and Quartz Fiber filters,

Microprocessor Based Electronic Control Module

Functions

* Flow rate metering and Control

* Barometric Pressure Measurement

* Volume Totalizer

* Temperature Measurement and Control

* Human Machine Interface through

  display and keyboard

* Real Time Clock & Timer

* Data transfer to PC or to data storage module

Programmable Real Time Clock / Timer

Accuracy :+/- 2 minutes / month   

Functions : date and time of Sampling, Automatic Start / Stop

 Digital Display

20 Characters X 4 lines LCD Module

Parameters Displayed (Instantaneous values )

  Flow Rate, Time Of Sampling  and Total Volume of Sampling

     Barometric Pressure Ambient Temperature and Filter Temperature

Flow Rate & Volume Measurements

Measuring Technique

Flow rate is inferred from the inlet air pressure, its temperature and  differential pressure across an orifice and corrected for actual operating conditions

Settable Range

16.67 LPM Ãâہ¡ÃƒÆ’‚±10% 

Accuracy

± 2% of reading

Flow Control System

Electronic

Flow Control Element

Flow Proportioning Valve

Accuracy of Control

Better than +/- 3% over 24 hours @ 16.67 LPM

Volume Measurement

Units : M3   Resolution : 0.01 M3

Totalisation and updating periodicity : 5 seconds from above flow rate

Computations,

recording,   reporting  and display on demand

Averaging : every 5 minutes

Reporting : 5 minutes average Out of specs by 5%

Computations: Flow average & Coefficient of variance (CV)

Ambient and Filter Temperature Measurement

Range & Resolution

00C to 500C with a resolution of 0.10 C

Temperature Control

Differential between Ambient and Filter Temperatures is maintained within Ãâہ¡ÃƒÆ’‚±50C by operation of a fan. 

Recording and Display

Average, Min. and Max value of both temperatures.  Maximum differential value and period of occurrence  of these Temperatures if it exceeds  50C

Barometric Pressure Measurement

Range & Resolution

600 to 800 mm Hg with a resolution1 mm Hg

Recording and Display

Average, Minimum and Maximum

 System Leak Rate Test

 Internal and External with Leak Rate Adaptor

Leak Rate

< 80 cc per minute

Reporting

Leak rate value

Data Transfer port

RS-232 serial port for data transfer to PC or memory module

Vacuum Pump

A diaphragm type pump with AC motor.

Power

230 V AC ± 10%, 50 Hz.( 110 V, 60 Hz Optional )

Construction

* Sampling Inlet, Size Separator and Impactors made of Aluminum with machining, heat- treatment, and surface coating as per EPA standards

* Suitable for Outdoor Use, powder coated metallic housing for Electronics etc.

* Over all Height: 2mtrs ( from ground to top of Sampling inlet) with a Tripod stand





 

 